7.4 Manual Calibration:
The HC1000 offers a quick manual calibration mode, allowing the user to
easily calibrate the meter.
7.4.1
Record the cell constant value, which is clearly marked on a label
attached to the electrode (e.g., K=1.08).
7.4.2
Ensure the Cal indicator is showing, and press and hold the CAL
key for 3 seconds to enter setup menu, the display shows “CELL” option.
7.4.3
Press ENTER key to confirm, the meter enters the cell constant
setting mode.
7.4.4
Press ▲ or ▼ key to select “USER” option.
7.4.5
Press ENTER key to confirm, the meter enters User-Defined mode.
Enter
MI
Enter