Now the method name is displayed inverse and the
test result is calculated with new calibration factor.
Delete User Calibration
This chapter only applies for the methods which can be user calibrated.
Select the required method.
Instead of zeroing the instrument, press [MODE],
[Shift] + [4] [6].
Confirm with [ ] key.
The display shows:
Press [Shift] + [1] keys to delete user calibration.
Press [Shift] + [0] keys to keep the valid user
calibration.
The instrument goes back to zero-query
automatically.
100
Chlorine T
0.02-6 mg/l Cl2
1.00 mg/l free Cl2
100
Chlorine T
0.02-6 mg/l Cl2
Prepare Zero
press ZERO
<User calibration>
100
Chlorine T
0.02-6 mg/l Cl2
1.00 mg/l free Cl2
Clear user calibration?
YES: 1, NO: 0