11.3 Conductivity Electrode Care and Maintenance
Ensure that the conductivity electrode is thoroughly washed with distilled
water after each use.
If there is a build-up of solids inside the measurement area of the cell,
these should be removed very carefully with a cotton bud soaked in
solvent, taking care not to touch the metal parts of the inner cell.
11.4 DO Probe Care and Maintenance
Always keep probe membrane moist.
If dissolved oxygen probe is not in use for long periods, unscrew
membrane cap and rinse the probe's cathode, anode and the membrane
with deionised water, then soak up residual water with filter paper, and re-
install the probe.
Cathode and Anode
Membrane cap