3.2 Getting started
Instructions for membrane filtration are covered in another section of the kit
manual.
The user should familiarise themselves fully with the set-up, control and
calibration of the incubator prior to carrying out tests for the first time.
If the incubator is connected to the docking station turn on using the switch on
the rear of the carry case. If connected directly to the power supply turn on the
power at the AC mains outlet.
The incubator screen will turn on.
The incubator may be in one of three operating modes on power up:
1: New run
2: Partial incubation cycle (Idle)
3: Partial incubation cycle (Run)
If the incubator is in mode 1 or 2 the screen shown in Fig 1.0 (see page 8) will be
displayed. If the incubator is in mode 3 the incubator will return to the last know
state and auto start the incubation cycle from where it was before the power
was removed. Mode 3 is primarily intended for unattended power failure
conditions.
If Fig 1.0 is displayed and the incubator was in mode 2 then the previous
incubation cycle can be restarted from where it was left by pressing restart. This
will be discussed later.
The screen will be displayed for 60 seconds before defaulting to the restart
condition. Pressing the right-hand switch will reset the incubator ready to set-up
and start a new incubation cycle and the screen will change to the one shown in
Fig 1.1 (see page 8).
The incubator has two independently controlled thermal chambers which can be
configured and calibrated independently. The left-hand switch primarily selects
the mode for the left-hand chamber and the right hand switch the mode for the
right-hand chamber. The centre switch will perform selections for each side
which will be indicated by an arrow pointing left or right < >.
Each screen and options will now be discussed in further detail.
Содержание WSL50 Pro
Страница 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Страница 4: ...Section 2 Kit Contents c c c c b a e f d ...
Страница 7: ...Section 3 ...
Страница 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Страница 10: ...6 0 Guarantee and Assistance 32 ...
Страница 12: ......
Страница 16: ......
Страница 21: ......
Страница 33: ...Section 4 ...
Страница 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Страница 50: ...Section 5 HydroTest Photometer HT1000 ...
Страница 86: ......
Страница 136: ......
Страница 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 193: ...NOTES ...