4.3 Maintenance
The wheels should be periodically cleaned. When dry add a drop of lubricating oil
to the wheel axles to prevent seizing.
All screws in the case are fitted with anti-vibration nuts or secured with adhesive.
However, after transporting the case in a vehicle, it is suggested to check the
integrity of the docking station and any screws to ensure no damage has
occurred and all electrical connections are secure prior to use.
After a period of time replacement of the rechargeable lead acid batteries may
be necessary. If so these must be replaced with batteries of the same type as
specified in section 5.1 and replaced as a pair.
Should any fault occur please contact the Trace2o® Technical Team who will be
more than happy to advise you.
5.0 Technical Specification
5.1 Docking station:
Power supply
Input (Max)
12-28V DC, 48W ( via supplied
Power supply unit or vehicle
charging lead only)
Output (Max)
12-28V DC, fused 24W
Fuse
2A Type T
Environmental
Operating temperature range
15 – 50°C
Storage temperature
0 – 70°C
Protection
Water ingress (Closed)
IP68 (1hr, 0.2m)
Buoyancy
47Kg
Batteries
Panasonic UP-VW1245P1
Rechargeable Lead Acid Battery 12V,
7.8Ah
Содержание WSL50 Pro
Страница 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Страница 4: ...Section 2 Kit Contents c c c c b a e f d ...
Страница 7: ...Section 3 ...
Страница 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Страница 10: ...6 0 Guarantee and Assistance 32 ...
Страница 12: ......
Страница 16: ......
Страница 21: ......
Страница 33: ...Section 4 ...
Страница 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Страница 50: ...Section 5 HydroTest Photometer HT1000 ...
Страница 86: ......
Страница 136: ......
Страница 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 193: ...NOTES ...