3.3.5 44°C Program (Fig1.3)
This mode is the pre-programmed default 44°C incubation cycle. In this mode the
incubator will run a 24 hour incubation cycle at 44°C. The timer can be altered to
any setting between 0 and 24 hours. Adjusting the incubation time is described
in section 3.3.8.
Pressing and holding the middle switch for 5 seconds will start the incubation
cycle and the screen will change to the following:
The asterisk preceding the temperature will switch on and off indicating the
status of the heater.
To stop or pause the incubation cycle press and hold the middle switch again for
5 seconds. The display will show the following:
Pressing and holding the middle switch will restart the cycle or pressing the left
switch will return to the menu function screen.
20.9°C 20.9°C
->44°C MEM
OFF
*20.9°C 20.9°C
24:00 RUN! 24:00
*43.9°C 20.9°C
18:00 MEM 24:00