4.1 Setup Menu
When in setup mode, the HC1000 displays an integrated setup menu that
allows you customise to the operation of the meter to meet your
measurement requirements. In the different modes, the meter will show the
corresponding options. For general options, the change will be applied to all
modes.
pH MODE:
DISPLAY
DESCRIPTION
OPTION
S
DESCRIPTION
DEFAUL
T
Select from
common pH
buffer standard
groups for
calibration.
USA (1.68, 4.01,
7.00, 10.01,
12.45)
●
NIST (1.68, 4.01,
6.96, 9.18,
12.45)
DIN (1.09, 4.65,
6.79, 9.23,
12.75)
User-Defined
Buffer
Select the
number of
calibration points
to use when
calibrating the
meter
1 point
2 points
3 points
●
4 points
5 points
Set the resolution
for pH
measurement.
0.001pH
●
0.01pH
0.1pH
Set the
temperature unit.
Degrees Celsius
●
Degrees
Fahrenheit
Содержание WSL50 Pro
Страница 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Страница 4: ...Section 2 Kit Contents c c c c b a e f d ...
Страница 7: ...Section 3 ...
Страница 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Страница 10: ...6 0 Guarantee and Assistance 32 ...
Страница 12: ......
Страница 16: ......
Страница 21: ......
Страница 33: ...Section 4 ...
Страница 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Страница 50: ...Section 5 HydroTest Photometer HT1000 ...
Страница 86: ......
Страница 136: ......
Страница 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 193: ...NOTES ...