4. Within 5 seconds, the pump will respond with a number of confirmation beeps equal to the
number of times you pressed the audio bolus button.
Note: Do NOT press any of the function buttons at this time unless you wish to cancel delivery.
5. Within 5 seconds, the pump will beep twice to “ask” you to confirm that you wish to activate
delivery and the Confirm screen is displayed.
6. Within 5 seconds, press the button again to activate delivery. The pump will beep twice to
confirm your delivery command. The Delivering Bolus screen is displayed and your pump will
beep once to signal the start of delivery and once to signal end of delivery (if you turned on
Normal Bolus Sounds in Setup).
7. The pump will again sound the number of beeps to indicate the step size you’ve chosen and
to signal the Audio Bolus sequence is finished.
If you wish to cancel the Audio Bolus, press any button. The
bolus will be canceled (as in the case of a Normal bolus).
If you cancel a bolus delivery after you’ve activated it, the screen at
right will be displayed. See Chapter 14, Safety System and Alarms.
If during a bolus delivery, your low cartridge level is reached,
the pump will not display the warning until after the bolus is
completed. So you could possibly have less insulin available
than your low cartridge setting.
Example: Your low cartridge warning is set to 10 (units). You have 20 units of insulin
remaining and you deliver a 15 unit bolus. The Low cartridge warning appears after bolus is
completed and you have only 5 units left – not 10 units. The Wake up screen will show the
remaining insulin amount.
The maximum number of Audio Bolus button presses is 20. Therefore, if you’ve set the step
size to 0.1 U, the maximum audio bolus amount is 2 U. If you’ve set the step size at 0.5 U,
the maximum audio bolus amount is 10 U and if your step size is 1.0 U, the maximum audio
bolus amount is 20 units. With a 5.0 U step size, the maximum cannot be greater than 35
units, which is the maximum amount for any type of bolus.
ezBolus™
If you do not have your Audio Bolus feature turned on, the button on the end of the pump will
function as a short cut to the Normal Bolus screen.
1. Press the button on the end once. The Normal Bolus screen is displayed. Program a
Normal Bolus as usual.
Содержание IR 1200
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 22: ...Page 22...
Страница 32: ...Page 32...
Страница 50: ...Page 50...
Страница 56: ...Page 56...
Страница 68: ...Page 68...
Страница 70: ...Page 70...
Страница 82: ...Page 82...
Страница 99: ...Page 99...