Page 35
Chapter 7 Using Basal Program Features
7
You can program your IR 1200 to display either 1 or 4 basal program options. Having more
than one pre-set basal program makes it easy for you to switch based on your weekday needs
versus weekend needs, accommodate for shift work and menstruation. If you are new to
pumping, your healthcare team may suggest you first become comfortable with one program
before using multiple basal programs. The factory default displays one basal program and the
temporary basal program. To display multiple basal options, s
ee Chapter 11, Advanced
Features.
Setting a Basal Program
Each basal program can be set with up to 12 different basal rates (doses) in a 24 hour period.
These 12 start times can be set to accommodate your changing basal needs throughout the
day. For example, your body may need more insulin in the early morning to compensate for the
“dawn phenomenon.” You can program time segments to begin at any hour or half hour.
Note: The Up/Down buttons will move you through fields when not in edit mode. When in Edit
mode, the Up/Down buttons will change the value of the field. If the cursor is flashing, that
means you can edit the entry. Use the OK/Enter button to start/stop Edit mode (flashing
cursor).
Use the chart in Chapter 18, My Information, to record and enter your basal program times and
doses as recommended by your healthcare team.
1. From the Main Menu, select Basal.
2. The Basal Menu displays the following:
•
Total basal insulin programmed for the 24 hour period
•
Temp (if you wish to program a Temporary Basal rate)
•
The active basal program, designated by number and by name, as well as an “A” to
indicate the active program. (If you have activated multiple basal programs in Setup
Advanced, all 4 basal program options will be displayed as shown on screen example
below, right.)
3. Select the desired program. Press OK/Enter.
Note: You do not have to Suspend the pump to edit an active program. When you select Edit
from the Basal Menu, your IR 1200 automatically suspends delivery. When you exit the Edit
mode, the active program delivery automatically resumes.
Содержание IR 1200
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 22: ...Page 22...
Страница 32: ...Page 32...
Страница 50: ...Page 50...
Страница 56: ...Page 56...
Страница 68: ...Page 68...
Страница 70: ...Page 70...
Страница 82: ...Page 82...
Страница 99: ...Page 99...