6. Remove cap from the filled cartridge tip. To avoid insulin spillage and introduction of air in
the cartridge, it should never be filled beyond the 2.0 mL mark. The plunger is properly
positioned for maximum fill when the black o-ring nearest the plunger tip is centered on the 2.0
mL mark. Attach infusion set Luer connector to cartridge tip using clockwise motion until finger
tight
and then twist another quarter of a turn.
Be sure to twist the Luer connector an extra quarter of a turn to ensure a secure
connection.
7. Put cartridge/tubing assembly aside.
Check for leaks, cracks or damage each time you change your cartridge and infusion set. To
avoid leakage, be sure to tighten the Luer connection securely. You can check for moisture
periodically by wrapping a tissue around the Luer connection between the cartridge and
infusion set.
When handling the cartridge, take care not to move the plunger from side to side in the
cartridge body. Maintaining straight alignment of the plunger keeps the o-rings properly
seated, which minimizes the possibility of introducing air into the cartridge and will prevent
insulin spillage.
Changing the Cartridge
1. Disconnect infusion set from your body.
2. Unscrew the cartridge cap, leaving tubing connected to the cartridge.
3. With the tubing connected to the cartridge, pull cartridge straight out of the pump.
4. Disconnect tubing from cartridge. Proceed with filling the new cartridge as outlined in this
chapter.
Connect infusion set to
cartridge here
Содержание IR 1200
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 22: ...Page 22...
Страница 32: ...Page 32...
Страница 50: ...Page 50...
Страница 56: ...Page 56...
Страница 68: ...Page 68...
Страница 70: ...Page 70...
Страница 82: ...Page 82...
Страница 99: ...Page 99...