Page 20
4
Audio Bolus/ezBolus
This button allows you to program a bolus discreetly, without looking at the pump, by
using audible tones to confirm programming and delivery*. If you choose not to activate the
Audio Bolus feature, this button provides a shortcut to the Normal Bolus screen. *When you
first use the Audio Bolus feature, you should always look at the screen to confirm correct
programming until you are comfortable with using audio feedback to program a bolus.
Battery Cap/Vent
This cap unscrews easily with a coin or the battery cap tool to replace and secure your battery.
There is a colored (blue or yellow) o-ring around the cap, which prevents water from entering.
The battery cap also is equipped with a vent built in to allow air to enter the pump to maintain
pressurization but prevent water from entering.
Primary Vent
This vent is part of the redundant vent safety system,
which allows air inside the pump to maintain equalized
pressure but prevents water from getting inside.
Under no circumstances should you introduce
any kind of sharp object into the vent
openings to clean them. Doing so will compromise
the pump’s waterproof capabilities. If at any
time you suspect the vent openings are clogged,
replace the battery cap or call Animas for questions about the Primary Vent.
Cartridge Compartment Cap
This cap secures your cartridge and infusion set in the pump.
IR Window for Download
The IR window is framed in blue. This is the infrared communication window used for
downloading* your pump data. *ezManager Plus software and accessory kit sold separately.
Outside the U.S., check with your local Animas Distributor.
Sounds
Your IR 1200 allows you to customize the volume level or use the vibrate function to notify you
of warnings and alarms and to confirm certain deliveries.
Tamper Resistant Feature
You can use the tamper resistant feature to prevent accidental button pressing. Simply wake up
the pump and press the Up and Down arrow buttons at the same time. This locks the pump
programming. To unlock the pump, wake up the pump so the screen reads “Locked”
and press
the up and down buttons at the same time again.
Cartridge Cap
Primary Vent
Battery Cap/Vent
2003-05
S/N: 49-00015-10
|||||||||||||||||||||||||| |||||
IR Window
Содержание IR 1200
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 22: ...Page 22...
Страница 32: ...Page 32...
Страница 50: ...Page 50...
Страница 56: ...Page 56...
Страница 68: ...Page 68...
Страница 70: ...Page 70...
Страница 82: ...Page 82...
Страница 99: ...Page 99...