Page 61
12
3. The Actual field is highlighted and flashing to indicate Edit mode. Use the Up/Down arrows
to enter your BG value. Press OK/Enter.
4. The Target field is highlighted.
a. If the target is correct, scroll to the Insulin Sensitivity Factor (ISF) field and go to Step 5.
Consult your healthcare team for your target BG values.
b. If the target is incorrect, press OK/Enter to activate Edit Mode. Use the Up/Down arrows
to enter the correct value. Press OK/Enter.
5. The Insulin Sensitivity Factor (ISF) field is highlighted.
a. If the ISF is correct, scroll down to Show Result. Press OK/Enter and go to Step 6.
b. If the ISF is incorrect, press OK/Enter to activate Edit Mode. Use the Up/Down arrows to
enter the correct value. Consult your healthcare team for your ISF. Press OK/Enter to exit
Edit mode. Show Result is highlighted. Press OK/Enter.
6. The Bolus Calc screen is displayed and the Bolus amount field is highlighted and flashing.
7. Enter the Bolus amount. Go is highlighted. Press OK/Enter to deliver or scroll to Type field
to select Combo bolus.
See this chapter, Combo Bolus.