Page 29
Priming the Pump and Infusion Set
5
These steps must be followed in order.
1.
Make sure you are disconnected from the pump.
2. From Main Menu, select Prime/Rewind.
3. On PrimeSmart menu, Rewind is highlighted. Press OK/Enter. The Rewind Motor screen is
displayed.
4. Scroll up to Go Rewind. Press OK/Enter. The pump will rewind the piston rod.
5. When the rewind action is complete, the Rewind Complete screen is displayed. The pump
will beep once to let you know the rewind is complete.
Note: If using a partially filled cartridge, you can select Stop during the
Rewind Active function to stop the rewind at the position desired. Periodically, the pump is
required to do a Full Rewind and will not offer the option of selecting the Stop position. A Full
Rewind is always required when a battery is inserted.
6. Insert your filled cartridge.
7. Secure cartridge compartment cap to pump by turning in a clockwise
motion until finger tight but
do not over tighten.
Note: If screen display has timed out while loading your cartridge, select
Prime/Rewind from Main Menu and highlight Load Cart from
Prime/Rewind Menu. Press OK/Enter to display Rewind Complete
screen. Continue with Step 8.