Page 62
12
ezBG
This feature allows you to enter your BG reading and your IR 1200 will calculate a correction
bolus. If the Insulin on Board (IOB) feature is activated, your IR 1200 will calculate a reduced
dose for your review.
1. From the Bolus Menu, select ezBG. Press the OK/Enter button.
2. The Actual field will be highlighted and flashing to indicate Edit mode. Use the Up/Down
arrows to enter your actual BG reading. Press the OK/Enter button to confirm the entry and exit
Edit mode.
3. Make sure the target amount is correct. Your healthcare team will give you your target BG
level. If you need to edit this field, press OK/Enter to activate Edit mode. Use Up/Down arrows
to change target. Press OK/Enter to confirm target number and to exit Edit Mode.
4. The Insulin Sensitivity Factor (ISF) field is highlighted. Check to be sure it is correct. Your
healthcare team will give you your ISF. If you need to edit this field, press OK/Enter to activate
Edit mode. Use Up/Down arrows to change target. Press OK/Enter to confirm ISF and to exit
Edit mode.