Chapter 14 Safety System and Alarms
14
Alerts, Warnings and Alarms
Your Animas pump has a progressive
Warnings and Alarms safety system.
This means if
you do not confirm the Warning or Alarm,
it will progress to the sweep alarm with vibrate
within one hour.
At the high volume stage, if you do not confirm the Warning or Alarm, the
sweep alarm will begin and will not stop until appropriate action is taken.
Alerts are automatically displayed to remind you of a function that you’ve set or a condition that
exists.
Warnings are triggered for a variety of reasons. They require you to confirm the warning by
pressing OK/Enter and/or taking action to address the warning.
Alarms are triggered by several conditions. All require you to address the alarm by taking
appropriate action in order to clear the alarm condition.
Alert: Active Basal Program Empty
Cause
Empty basal program activated.
Effect
No basal deliveries.
Message
Displayed once for 3 seconds and each time manually
awakened.
Action
None required but can confirm or select Basal Menu.
Beeps/Vib
User selected, one time and each time manually
awakened. No progression.
Alert: Temp Basal Minimum Rate
Cause
Negative Temp Basal activated.
Effect
Basal delivery will not go below 0.025 U/hr.
Message
Displayed once for 3 seconds.
Action None
required.
Beeps/Vib
User selected, one time. No progression.
Содержание IR 1200
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 22: ...Page 22...
Страница 32: ...Page 32...
Страница 50: ...Page 50...
Страница 56: ...Page 56...
Страница 68: ...Page 68...
Страница 70: ...Page 70...
Страница 82: ...Page 82...
Страница 99: ...Page 99...