Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 61 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
The table below lists the Read Input Registers.
Function
Register
Length
Comment
Read Input Register (04)
30001
2
Sensor Unit 1 live reading
Read Input Register (04)
30003
2
Sensor Unit 2 live reading
Read Input Register (04)
30005
2
Sensor Unit 3 live reading
Read Input Register (04)
30007
2
Sensor Unit 4 live reading
Read Input Register (04)
30009
1
Sensor Unit 1 measurement display units
Read Input Register (04)
30010
1
Sensor Unit 2 measurement display units
Read Input Register (04)
30011
1
Sensor Unit 3 measurement display units
Read Input Register (04)
30012
1
Sensor Unit 4 measurement display units
Read Input Register (04)
30013
4
Sensor Unit 1 short description
Read Input Register (04)
30017
4
Sensor Unit 2 short description
Read Input Register (04)
30021
4
Sensor Unit 3 short description
Read Input Register (04)
30025
4
Sensor Unit 4 short description
Read Input Register (04)
30029
4
DOM status flags (see below)
The table below lists the DOM status flags.
Register value (HEX)
DOM status
0 (00)
No fault
1 (01)
DOM firmware corrupt
2, 4 or 6 (02, 04 or 06)
No / corrupt Ax60+ comms
8 (08)
No Ax60 configuration
16 (10)
Corrupt Modbus comms
32 (20)
DAC convertor failed
64 (40)
Corrupt DOM configuration
128 (80)
Corrupt DAC configuration