11.2
Data Output Module fault indication
The DOM indicates its internal operating condition in one of three ways:
1)
By its Status and Fault LEDs (see below).
2)
By setting all of the 4-20mA channels to indicate a fault condition.
3)
By setting Modbus registers which can be read via the Modbus interface (See Appendix D for more
information).
11.2.1
Status and Fault LEDs
The table below shows the combination of LEDs which indicate a specific operating condition:
Status LED
Fault LED
Meaning
Flashing
Off
Normal operation
On
Flashing
Critical Fault (see below)
Off
Flashing
Fault (see below)
On
On
Warm-up (30 seconds)
A Critical Fault is:
•
An error has occurred with the internal storage memory of the unit and needs replacing.
A Fault can be one of the following:
•
The unit is not receiving data from the Ax60+ system. Check the cabling.
•
The unit has not received valid data from the Ax60+ system. Check the cabling and / or
power cycle the system.
•
If connected to a BMS, an invalid message has been received over the Modbus interface.
Check the cabling.
•
An internal error has occurred with the unit and will it have to be replaced.