Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 12 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.1
Read-only menus
The Show config option enables you to view read-only information about the system configuration. It displays
information on Sensor connections and relay configuration.
NOTE:
NOTE: THE Show config OPTION PROVIDES READ-ONLY INFORMATION THAT CAN
NOT BE CHANGED. IF YOU WANT TO SET UP OR CONFIGURE THE Ax60+, USE THE Top-level Menus,
Central Config OPTION.
4.1.1
Show configuration (Displayed as Show config)
1]
To display the Top-level Menu Show config option, press and hold down the Cycle
buttons for at least 6 seconds.
2]
Press Accept/Test to display the Show Config? option
3]
Press Cycle to cycle through each of the config options
NOTE:
TO RETURN TO THE SYSTEM STATUS SCREEN AT ANY TIME PRESS Cancel. THE Cancel
BUTTON REDISPLAYS THE PREVIOUS MENUS, ONE AT A TIME.