9.7
Alarm set-points (Displayed as Alrm set-pnts)
CAUTION: ALARMS ARE LATCHING. THIS MEANS THE SENSORS ALWAYS REMEMBER THEIR
CURRENT ALARM STATE, EVEN AFTER A POWER OUTAGE. FOR EXAMPLE, IF A SENSOR IS EXPOSED
TO 3% CO
2
BUT THE ALARM IS NOT ACKNOWLEDGED AND THE POWER IS REMOVED, WHEN
POWER IS RESTORED IT GOES STRAIGHT INTO ALARM.
CAUTION: DEFAULT ALARM LEVELS SHOULD BE CHANGED ONLY BY AN AUTHORISED AX60+
INSTALLER OR SERVICE ENGINEER.
NOTE:
ALARM LEVELS ARE STORED IN EACH SENSOR UNIT, THESE WILL NEED TO BE SET AT
SENSOR INSTALLATION IF THE REQUIRED LEVELS DIFFER FROM DEFAULT.
1]
Press and hold Cycle for at least 6 seconds. Then press Cycle six times to display
Sensor Config.
2]
Standard increments for alarm setpoints is factory set at 0.10 %, however, by pressing and
holding the Cycle button this increases the increment value x 10 (1.00 %)
3]
Press Cycle six times to display the Alrm set-pnts screen
4]
Press Accept/Test to display the Select sensor screen, press Cycle to toggle to chosen sensor,
then press Accept/Test to display the Select alarm screen
5]
Press Cycle to toggle through the alarms, then press Accept/Test to display the Set alarm AL?
screen (Where ? could be AL1, AL2, AL3 or AL4)
6]
Press Cycle repeatedly to enter the alarm value you want in 0.10% increments, or press and
hold Cycle to increase the value at 1.00% increments.
7]
To reset the value and start again, repeatedly press Cycle or press and hold Cycle until you
exceed 6.00 %, this will take the value back to 0.00 % and you can cycle through again until
you reach the required value.
8]
Press Accept/Test to accept the alarm set point.