Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 43 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
8.4.8
Relay 2 on alarms (Displayed as Rly 2 on alarms)
1]
Press and hold Cycle for at least 6 seconds. Then press Cycle five times to display
Central Config.
2]
Press Accept/Test, then press Cycle eight times, press Accept/Test to display the Rly 2 on
alarms option.
3]
Press Accept/Test to display the Select sensor screen, press Cycle to toggle to chosen sensor,
then press Accept/Test to display the Select alarm screen
4]
Press Cycle to toggle to chosen alarm, press Accept/Test to display the Set on Alarm option,
press Cycle to toggle Y or N, then press Accept/Test to confirm.
5]
Press Cancel repeatedly to return to system status screen