Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 60 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
16.2
MODBUS register map
The following table shows which Modbus registers hold additional information about the operation of the DOM and
Ax60+ system.
Data is sent in big-endian format on the wire
.
Important restriction: the DOM only supports the requesting of one item at-a-time. For example, to request the contents
of Read Input register 1 (Sensor Unit 1’
s gas reading) the request message will be for only two registers (1 & 2). If a
request is made for multiple items, then exception code “Illegal Function” will be returned.
The table below lists the Read Input Status.
Function
Register
Length
Comment
Read Input Status (02)
10001
1
Sensor Unit 1 fault condition
Read Input Status (02)
10002
1
Sensor Unit 2 fault condition
Read Input Status (02)
10003
1
Sensor Unit 3 fault condition
Read Input Status (02)
10004
1
Sensor Unit 4 fault condition
Read Input Status (02)
10005
1
Sensor Unit 1 alarm level 1 active
Read Input Status (02)
10006
1
Sensor Unit 2 alarm level 1 active
Read Input Status (02)
10007
1
Sensor Unit 3 alarm level 1 active
Read Input Status (02)
10008
1
Sensor Unit 4 alarm level 1 active
Read Input Status (02)
10009
1
Sensor Unit 1 alarm level 2 active
Read Input Status (02)
10010
1
Sensor Unit 2 alarm level 2 active
Read Input Status (02)
10011
1
Sensor Unit 3 alarm level 2 active
Read Input Status (02)
10012
1
Sensor Unit 4 alarm level 2 active
Read Input Status (02)
10013
1
Sensor Unit 1 alarm level 3 active
Read Input Status (02)
10014
1
Sensor Unit 2 alarm level 3 active
Read Input Status (02)
10015
1
Sensor Unit 3 alarm level 3 active
Read Input Status (02)
10016
1
Sensor Unit 4 alarm level 3 active
Read Input Status (02)
10017
1
Sensor Unit 1 alarm level 4 active
Read Input Status (02)
10018
1
Sensor Unit 2 alarm level 4 active
Read Input Status (02)
10019
1
Sensor Unit 3 alarm level 4 active
Read Input Status (02)
10020
1
Sensor Unit 4 alarm level 4 active
Read Input Status (02)
10021
1
DOM fault condition