Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 29 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
6
Calibrating Ax60+ Kiosk sensors
6.1
CO
2
sensor calibration proof test
The Ax60+ CO
2
Sensors can be checked to ensure that they accurately measure carbon dioxide and trigger the alarms
correctly. Analox can supply the following check kit:
Ax60+ CO
2
Check Kit comprising:
Part number
Description
Contact Analox or source locally
110-litre 1.6% carbon dioxide, balance nitrogen (low alarm check gas)
Contact Analox or source locally
110-litre 3.2% carbon dioxide, balance nitrogen (high alarm check gas)
Contact Analox or source locally
0.5-litre/minute regulator to fit 110-litre bottle
1817-5000
2-metre neoprene tubing
8000-0910A
Push-in flow adaptor
Checking CO
2
Sensors is an easy task that does not require software input. This procedure is suitable for the Ax60+ Kiosk
option only. Follow the procedure below:
NOTE:
IF THE OPTIONAL SENSOR SPLASHGUARD IS FITTED, REMOVE IT BEFORE CHECKING
AND/OR CALIBRATING THE SENSORS.
1]
Ensure the Ax60+/Ax60+ Kiosk is powered on and operating normally.
2]
Fit the neoprene tubing (1817-5000) to the flow adaptor (8000-0910A).
3]
Fit the flow adaptor into the aperture on the CO2 Sensor (see below, left).
Step [3] Fit the flow adaptor into
the aperture on the CO
2
Sensor
Step [4] Fit the regulator to the 1.6% CO
2
gas bottle
4]
Fit the regulator to the 1.6% CO2 gas bottle (see above, right).
Содержание Ax60+ Multi-Gas