Ax60+ Multi-Gas
Service Manual
Int. Approved
Document ref: P0159-803-11
September 2019
Page 18 of 62
Copyright © 2015 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.4
Testing relays (Displayed as Test relays)
The Test relays option is used to switch Relays 1 and 2 to check they are functioning.
WARNING: DURING A RELAY TEST, ANY CONNECTED DEVICES OR CIRCUITS WILL OPERATE.
DISCONNECT AND ISOLATE ANY DEVICES OR CIRCUITS CONNECTED TO THE RELAYS BEFORE
TESTING THEM.
NOTE:
IF YOU FORGET TO SWITCH OFF THE RELAYS DURING THE TEST THEY WILL
AUTOMATICALLY SWITCH OFF WHEN YOU EXIT THE MENU OR WHEN THE SOFTWARE TIMES OUT
AFTER 60 SECONDS.
NOTE:
RELAYS ARE NOT TRIGGERED WHEN CARRYING O
UT THE ‘ALARM
TEST’ FUNCTION.
1]
To display the Top-level Menu Test relays option, press and hold down the Cycle
buttons for at least 6 seconds. Then press the Cycle button three times.
2]
Press Accept/Test to go to the Activate Rly 1 screen, Press Accept/Test to activate the relay
3]
Press Cycle to go to Activate Rly 2 screen, Press Accept/Test to activate the relay
4]
Press Cycle to go to Rlys inactive screen, Press Accept/Test to deactivate the relays
NOTE: YOU WILL HEAR CLICKS FROM THE CENTRAL DISPLAY DURING TESTING OF RELAYS
5]
Press Cancel to return to Top-level Menu, Test relays