PN 9000-133-VE, Rev U
Communication Troubleshooting Checks
No Communications
Possible Cause
Corrective Action
Polarity reversed
Check the wiring for proper polarity. Review associated wiring
diagrams. Correct the wiring.
Address not set correctly
From the Display, check the Address setting.
On the
User Interface
board, check the dip switch settings.
Refer to “Setting the Analyzer Address” in this section for more
information.
Maximum distance be-
tween devices exceeded
If the distance between the Display Unit and the analyzer
(furthest analyzer in a daisy-chain configuration) exceeds 1219.2
meters (4000 ft), the devices must be moved closer together.
Broken connection
Check cables and wiring for damaged or broken connections.
Replace or fix the connections as needed.
Intermittent Communications
Possible Cause
Corrective Action
Poor wiring (i.e., lack of
shielded twisted pair)
Check the wiring to ensure it meets proper specifications for this
installation. Replace the wiring if necessary.
Termination enabled on
more than one unit (on
multi-unit systems)
Check the communication termination resistors (dip switch
SW-5
) on each analyzer in the daisy-chain. This switch is located
on the front of the
Sensor
board. Only the dip switch on the
last analyzer (farthest from the Display Unit should be Enabled.
Ensure all others are Disabled.
Termination not enabled
(on multi-unit systems)
Maximum distance be-
tween devices exceeded
If the distance between the Display Unit and the analyzer (fur-
thest analyzer in a daisy-chain configuration) is close to exceed-
ing, or does exceed, the maximum required distance (1219.2 me-
ters/4000 ft), the devices may need to be moved closer together.
2-44 | Thermox
®
WDG-V / VC / VCM / VM
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...