PN 9000-133-VE, Rev U
FURNACE
O-RING
CELL
HEX NUT
CELL HOUSING
WASHER
HOLD CELL HOUSING
USING 7/8" OR 1" WRENCH
(23mm or 26mm)
TURN HEX NUT COUNTER
CLOCKWISE USING A
7/8" WRENCH (23mm)
CELL CLIPS
CELL REPLACEMENT
Figure 3-3.
Cell replacement.
6. Place the Cell o-ring on the base of the Cell as shown in Figure 3-3.
7. Insert the new Cell into the Cell housing. Do not touch the bare Cell when
reinserting. Tighten the Cell into the Cell housing using the lower hex nut
(the upper hex nut is pre-tightened at the factory). The Cell o-ring provides
a seal for the system. Ensure the o-ring is properly aligned in its o-ring
groove and apply even pressure when tightening the Cell.
Maintenance and Troubleshooting
|
3-17