PN 9000-133-VE, Rev U
Process Pressure Checks (only if above 2 PSIG)
Problem/Possible Cause
Corrective Action
Process Pressure set
incorrectly
To check that you entered the process pressure correctly, select
Process Pressure
from the
Configuration
menu. Also ensure
that the calibration process pressure equals normal operating
process pressure. Calibration should only be performed under
these conditions for highest accuracy.
Cell Checks
Problem/Possible Cause
Corrective Action
If the Cell fails when you
first begin to use the
analyzer:
It is likely that there is a leak in the analyzer plumbing or an
improper calibration gas setup, and there is not a problem with
the Cell itself (see “Leak Check” in this chapter for help on how
to check for leaks; see “Calibration/Aspirator Air Setup Checks” in
this chapter for help on checking your calibration gas setup).
If the analyzer has been
operating for some time
and you feel the oxygen
reading is inaccurate:
First check by running a known calibration gas to verify the ana-
lyzer’s response. If the analyzer responds to the calibration gas
correctly, it indicates either leaking or plugged plumbing.
If your analyzer doesn’t
respond properly to the
calibration gas:
This may indicate a problem with the Cell. Before replacing the
Cell, check for leaks or plugged plumbing.
Maintenance and Troubleshooting
|
3-9
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...