PN 9000-133-VE, Rev U
WDG-V Mains Supply Connections
Do not run control unit AC mains supply wiring in the same conduit
with other AC line power wires. By keeping this wiring separated, you
prevent transient signals from reaching the control unit.
The WDG-V can operate using either 115 VAC, ±10 % or 230 VAC
, ±10 %
. There
is no power switch or circuit breaker on the analyzer, and it must be protected
by installing it on a circuit-protected line, maximum 15 amperes, with a switch
or circuit breaker in close proximity to the control unit and within easy reach of
an operator. Mark the switch or circuit breaker as the control unit disconnect-
ing device.
Mains supply connections to the control unit are:
• L – Line connection
• N – Neutral connection (USA)
• Chassis Stud – Equipment ground (protective conductor)
Use the 1/2" conduit entry hole in the WDG-V for AC mains supply wiring. Use
the chassis stud next to the 1/2" conduit entry hole for equipment ground
(protective conductor).
AC
(
L
) and (
N
) markings are provided by the terminal block for connection
of AC power. These markings are for reference purposes only, such as for use
on system wiring diagrams, etc. The system/product has or needs no specific
LINE
or
NEUTRAL
connection for any function, safety or otherwise. The (
N
)
terminal is not internally grounded, nor needs to be. The system will operate
normally regardless of what AC input terminal (
L
or
N
) the
AC Line
or
Neu-
tral
is connected to, or, if there is a Neutral used at all (i.e., 208 VAC US power
connection).
Installation and Start-Up
|
2-27
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...