PN 9000-133-VE, Rev U
Final Daisy Chain Connection Configuration Setting
When connecting multiple analyzers in a daisy chain configuration, enable the
communication termination resistor (Dip Switch
SW-5
) on the last analyzer
in the chain (farthest from the Display User Interface).
SW-5
is located on the
front of the
User Interface
board.
• SW-5 ON: Termination Enabled
• SW-5 OFF: Termination Disabled
The maximum distance between the AMEVision Display User Interface
and the analyzer is 1219.2 meters (4000 feet).
For instructions on how to work from the AMEVision Display User
Interface to configure and calibrate the analyzer, and to view analyzer
status readings, refer to the AMEVision Display User Interface
User
Manual
(AMETEK PN 9000-165-VE).
2-42 | Thermox
®
WDG-V / VC / VCM / VM
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...