PN 9000-133-VE, Rev U
Alarm Condition / Description / Corrective Action
Last Comb Calibration Failed
The last combustible calibration failed.
Trigger Conditions:
• Tank Empty.
• Bad solenoid.
• Insufficient time allowed for calibration.
• Bad detector.
Corrective Action:
• Check the Calibration tank.
• Check the solenoid for proper operation.
• Restart the Calibration and allow sufficient time for it to complete its cycle.
• Check the Combustibles Detector for proper operation.
Last Flow Calibration Failed
The last Flow Sensor calibration failed.
Trigger Conditions:
• Plugged Sample System.
• Faulty O
2
Cell.
Corrective Action:
• Check the Sample System (and Inlet Probe) for plugging problems. Refer to “Plugged Plumbing Check” for
information on how to check for plugging problems.
• Check the O
2
Cell for proper operation.
Last O
2
Span Calibration Failed
The last oxygen Span Calibration failed.
Trigger Conditions:
• Span Calibration gas tank is empty.
• Elapsed Span Calibration time too short.
• Faulty Calibration gas solenoid.
• Faulty O
2
Cell.
Corrective Action:
• Check the Span Calibration gas tank. Replace if necessary.
• Allow sufficient time to allow the Span Calibration gas to flow through the O
2
Sensor and to stabilize. Refer to
“Calibration/Aspirator Air Setup Checks” in this chapter for additional information.
• Check the Calibration gas solenoid for proper operation.
• Check the O
2
Cell for proper operation.
Maintenance and Troubleshooting
|
3-39
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...