PN 9000-133-VE, Rev U
Temperature Sensors
There are two tightly controlled temperature zones in the WDG-V. The Oxygen
Cell Temperature and the combustible block temperature (WDG-VCM units
only). (On Oxygen-only units the box temperature is controlled with an RTD
mounted on the aspirator block.)
Component
Function
Cell Temperature
The Oxygen cell is controlled to 680 °C. The cell temperature is
measured by dual, Type K Thermocouples. Dual Thermocouples
are used for self check of the cell temperature. The Thermocou-
ples are collocated in a single assembly. There is no distinction
between the two Thermocouples as they are effectively at the
same location in the assembly.
Combustibles/
Box Temperature
The combustible block/box temperature is controlled to 225 °C.
This temperature is measured by 100 ohm platinum RTD. The
RTD is embedded in the combustible block. On Oxygen-only
units, the RTD is mounted on the aspirator block. All internal
plumbing components are maintained above 200 °C.
Heaters
There are two main heaters in the WDG-V, the cell heater and box heater. In
WDG-VCM units, an additional heater is used for fine temperature control of
the combustible block.
Component
Function
Cell Heater
The Oxygen Cell heater is a 300 W coil heater encompassing the
cell housing and providing precision heat to the oxygen cell.
Box Heater
The box heater is a 350 W cartridge heater located in the alu-
minum heater block at the bottom of the aspirator. This heater
provides the heat for the enclosure to maintain all plumbing
components above 200 °C.
The combustible heater is a 50 W heater used for precise tem-
perature control of the combustible block.
Overview
|
1-9
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...