PN 9000-133-VE, Rev U
Warnings
Alarm Condition / Description / Corrective Action
Analog Output 1 Out Of Range
The analog output value is saturated (high or low).
Trigger Conditions:
• Incorrect range setting.
Corrective Action:
From the User Interface or PC Configurator Software:
• Check the range configured for Analog Output 1. Change it to the required range if necessary.
Analog Output 2 Out Of Range
The analog output value is saturated (high or low).
Trigger Conditions:
• Incorrect range setting.
Corrective Action:
From the User Interface or PC Configurator Software:
• Check the range configured for Analog Output 2. Change it to the required range if necessary.
Analog Output 3 Out Of Range
The analog output value is saturated (high or low).
Trigger Conditions:
• Incorrect range setting.
Corrective Action:
From the User Interface or PC Configurator Software:
• Check the range configured for Analog Output 3. Change it to the required range if necessary.
Cell Life Nearing Its End
The ZrO
2
Cell is nearing its end of life. This is based on calibration history.
Trigger Conditions:
• Low Cell output.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Replace the ZrO
2
Cell.
Contact AMETEK to verify operation and to order a new Cell.
Maintenance and Troubleshooting
|
3-37
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...