PN 9000-133-VE, Rev U
Alarm Condition / Description / Corrective Action
Flow Sensor Failure
The Flow Sensor failed (open).
Trigger Conditions:
• Faulty Flow Sensor.
• Faulty Flow Sensor circuit.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Check the Flow Sensor for proper operation.
• Check the Flow Sensor for a faulty circuit.
Low Sample Flow
Low sample flow detected.
Trigger Conditions:
• This condition may be triggered upon analyzer power-up while the analyzer is warming up, but will clear after
the analyzer has properly warmed up to its normal operating temperature.
• Loss of Aspirator Air.
• Plugged Inlet Probe.
• Plugged Sample System.
Corrective Action:
• Check the Aspirator Air line for damage.
• Check the Aspirator Air supply to ensure it is adequate.
Take appropriate safety precautions, power down the Analyzer, and:
• Check the Sample System and Inlet Probe for plugging problems. Refer to “Plugged Plumbing Check” for infor-
mation on how to check for plugging problems.
Methane Detector Open
(applicable only to WDG-VM Analyzers)
Methane Detector is open or not connected.
Trigger Conditions:
• Bad Methane Detector.
• Bad electronics or wiring.
Corrective Action:
Take appropriate safety precautions, open the Analyzer door, and:
• Check the Methane Detector for an open circuit.
• Check the operation of the electronics and check the wiring for damage or improper connections.
• Check the Methane Detector for proper operation. Replace the Detector if necessary.
Contact AMETEK to
verify operation before removing the Detector.
Maintenance and Troubleshooting
|
3-35
Содержание WDG-V
Страница 1: ...WDG V WDG VC WDG VCM WDG VM Combustion Analyzer USER MANUAL PN 9000 133 VE Rev U 5000 017 PI rA...
Страница 12: ...PN 9000 133 VE Rev U This page intentionally left blank xii Thermox WDG V VC VCM VM...
Страница 60: ...PN 9000 133 VE Rev U Figure 2 16 RCU Connections 2 34 Thermox WDG V VC VCM VM...
Страница 93: ...PN 9000 133 VE Rev U Figure 3 4 Thermocouple replacement Maintenance and Troubleshooting 3 19...
Страница 97: ...PN 9000 133 VE Rev U Figure 3 6 Flow sensor replacement Maintenance and Troubleshooting 3 23...
Страница 101: ...PN 9000 133 VE Rev U Figure 3 9 Combustibles Block Heater Replacement Maintenance and Troubleshooting 3 27...
Страница 103: ...PN 9000 133 VE Rev U Figure 3 10 Sensor board removal Maintenance and Troubleshooting 3 29...
Страница 124: ...PN 9000 133 VE Rev U This page intentionally left blank 4 10 Thermox WDG V VC VCM VM...