AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
9 / 80
1
"Test On"
2
Function keys "F1...F7
3
Display
4
Exit
5
Cursor keys "
" and "
→
"
6
Knob (Inc. / Dec)
7
Escape
8
CRO trigger output
5V
9
LED coupling 1 ph. version
10
EUT test supply L, N, PE 1ph version
11
LED coupling 3 ph. version
12
EUT test supply 3-phase version
13
Direct output
HV
–
COM
11 LED coupling
The LED Indicates the actual coupling mode for the 3-phase version.
12 EUT test supply
For single- or three-phase EUT the coupling/decoupling network is part of the generator. The EUT is powered via the safety
banana plugs at the front panel of the simulator
13 Direct HV and COM output
The HV and COM output are the designed for an external using of the impulse. This output is floating and is
used for external coupling/decoupling networks or magnetic field antenna.
Attention
The direct HV and COM output of the generator is located at the front panel of the instrument. It
is not allowed to connect these outputs to any other coupling/decoupling network than
manufactured by AMETEK CTS, e.g. the types CNV or CNI.
Before to connect any external networks to this output the operator must contact the
manufacturer. Any damages due to this matter are not covered by warranty.
The direct pulse output shall also not be used to connect the generator directly to any power
conducting lines.