11.3. USB Interface
USB interface “USB B” connector. For data transfer a USB interface is available. The internal RS 232
interface is converted to USB standard. Therefore the user must set the same Baud rate in the device and
control software.
Using the USB interface the user can have emc problems during burst tests Our experiences says, that
usually the computer USB port is disturbed by interference’s. Therefore a high quality USB cable (USB 2.0
standard) must be used.
•
K-USB
USB interface cable
High quality USB 2.0 interface cable for data transfer to the
computer.
Length: 3m connector type USB A
– USB B
USB cable setup
The USB cable must be above ground with a
distance of at least 10 cm. Otherwise the cable can
be an antenna for the common mode burst pulses
and will “collect” the interferences.
For filtering the following type of ferrites can be
used: Kitagawa TR-40-27-15 with 8 windings
AMETEK CTS deliver the ferrite under the name
FER-USB as an option
11.4. Optical Interface
For eliminate EMC interferences as earth loops, an optical link delivers a galvanic separation between
generator and computer for remote control.
OptoLink
- Fiber optic link and interface
with USB A connector
- Optical cable , length 3m
Содержание OCS 500 M6
Страница 76: ......
Страница 77: ...AMETEK CTS OCS 500 N6 series Manual for Operation V 5 22 77 80 12 2 OCS 500 N6 General Diagram ...