6.1.2.
Standard test routines
The user can select preprogrammed standard test routines.
Page 3
Damped Osc. 1MHz
IEC 61000-4-18
F1 : Common Mode
L-PE, N-PE L+N - PE
F2 : Differential Mode
L-N
F1
F2
F3
F4
F5
F6
F7
Page 4
Damped Osc. 1MHz
IEC 61000-4-18
F1 : Level 1
500 V
F2 : Level 2
1000 V
F3 : Level 3
2000 V
F7 : Manual
F1
F2
F3
F4
F5
F6
F7
The standard levels for common and differential mode can be changed in the setup menu.
6.1.2.1.
Level test
A selected level will step through all couplings in the selected Mode. The following parameters are used:
Defined per
standard
Voltage
selected level voltage
Repetition f
40/s (100 kHz)
400/s (1 MHz)
Polarity
each coupling
positive and negative
Coupling
Differential Mode
L-N
Common Mode
L-PE, N-PE, L+N-PE
Variable
parameters
Burst Time td
0.1s
– 99.9s
Repetition tr
0.0s
– 99.9s
Test duration
1 min.
Page 5
(standard test in stop mode)
(Running standard test)
Damped Oscillatory
IEC 61000-4-18 Level 3
Damped Oscillatory
IEC 61000-4-18 Level 3
V
=
2000V
Rep
= 400/s
V
=
2000V
Rep
= 400/s
Td
= 2s
Td
= 2s
Tr
= 2s
+ 2000 V
Tr
= 2s
T
= 1:00 min
L
L
Testtime
current coupling
0:00:21
START CHANGE
STOP Step
complete test time
0:02:21
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
- The Test time shows the elapsed time per actual test level as
well as the complete time of all triggered test levels within the
running test sequence.
- Pushing the function key F2 STEP will bring you into the next
iteration sequence.