AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
54 / 80
8.1.2.2.
Manual test routine
The manual test menu is only in the single phase equipment available.
Within this test routine all standard parameters can be easy set. A test cycle makes
10 impulses
within this
setting.
Manual standard test routine
Damped Osc. 1MHz IEC 61000-4-12
<- ->
O
O
Level 3
+
2000V 12 Ohm 0 dgr L-N
Testtime 0:00:45 h
Start
+
/- oV/A Imp L--PE N-PE L+N-PE Common mode menu
F1
F2
F3
F4
F5
F6
F7
Start
+
/- oU/A Imp L--N /
Differential mode menu
F1
F2
F3
F4
F5
F6
F7
Example:
- By pushing the cursor
➔
the test level will be increased/decreased to the next standard level.
- By turning the Inc. knob the test voltage V resp. the phase angle A will be adjusted continuously. The blinking
circle (
o
) shows which parameter can be changed.
Pressing the function „O V/A“ will change between both
parameters.
- Pressing the function keys the related function will be immediately activated.
- The displayed time will be resettled to zero after every new setting.
It is allowed to change all parameters during a running test