10.5. Verification Damped Oscillatory Wave 100 / 1000 kHz magnetic field
10.5.1. Verification Parameter
IEC 61000-4-10
(to MS 100 antenna)
Current measurement
Current
10.0A
±10%
Current
30.0A
±10%
Current
100.0A
±10%
H-Field measurement
Settings: 100A/m Antenna factor= 0.9
Current
111.1A
±10%
Oscillating frequency
100, 1000 kHz
±10%
Typical pulses, measured with a calibrated equipment
U set : +100A 100kHz Antenna factor = 0.9
U set : +100A 1000kHz Antenna factor = 0.9