6.3.
Menu Damped Osc. 1MHz EXT.
This menu appears only if an external coupling
network
CNV 508N4 is connected
via the CN
interface and recognized from the OCS 500N6 during
the power on routine. The menu structure is analogue
the tests with damped oscillatory for 100kHz and
1MHz.
CNV 5084N4
CNV 5084N4.1
Page 2
Damped Osc. 1MHz EXT.
IEC 60255-26
F1 : Quick start
F2 : Standard test routines
F3 : User test routines
F1
F2
F3
F4
F5
F6
F7
F1 Quick start
Easy and fast online-operation of the device.
F2 Standard test routines
The operator can select between various preprogrammed test routines as required in different standards
F3 User test routines
The operator can select between various preprogrammed test routines which helps to accelerate testing and
which are very helpful especially during design.
6.3.1. Coupling as per IEC 60255 with CNV 508 N4
The coupling network CNV 508 N4 offers the following coupling modes. The settings can be made on the
generator OCS 500N6 or with the software IEC.control.
Common Mode 1
L1+L2
– GND
L3+L4
– GND
L5+L6
– GND
L7+L8
– GND
All not tested pairs are not coupled; floating
Common Mode 2
L1+L2
– GND
(L3+L4+L5+L6+L7+L8) via Koppel C an GND
L3+L4
– GND
(L1+L2+L5+L6+L7+L8) via Koppel C an GND
L5+L6
– GND
(L1+L2+L3+L4+L7+L8) via Koppel C an GND
L7+L8
– GND
(L1+L2+L3+L4+L5+L6) via Koppel C an GND
All not tested pairs are connected via a coupling capacitor to generator GND
Differential Mode
L1 - L2
L3 - L4
L5 - L6
L7 - L8
Coupling between the lines of the selected pair X