AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
67 / 80
10.4.3. Verification Parameter fast damped oscillatory wave
This phenomenon is representative of switching of isolators in HV / MV open-air stations, and is particularly
related to the switching of HV bus-bars, as well as of background disturbances in industrial plants.
Verification as per IEC 61000-4-18:
Open circuit
V peak
250V
– 4000V
±10 %
Rise time first peak T1 5ns
±20 %
Oscillating frequency
3 MHz, 10 MHz, 30 MHz ±10 %
Decaying of peak 5
> 50% of peak 1
Decaying of peak 10
< 50% of peak 1
Repetition rate
5000/s
±10 %
Burst duration
3 MHz 50 ms
±20 %
10 MHz 15 ms
±20 %
30 MHz
5 ms
±20 %
Output Impedance
50
Ω
±30 %
Short circuit
Rise time first peak T1 3 MHz < 330 ns
10 MHz < 100 ns
30 MHz
< 33 ns
Osc. current frequency 3 MHz, 10 MHz, 30 MHz ±30 %
Decaying of peak 5
> 25 % of peak 1
Decaying of peak 10
< 25 % of peak 1
I peak
5 A to 80 A
± 20 %