AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
44 / 80
7.
Fast Damped Oscillatory 3 MHz, 10 MHz or 30 MHz as per IEC 61000-4-18
This phenomenon is representative of switching of isolators in HV / MV open-air stations, and is particularly
related to the switching of HV bus-bars, as well as of background disturbances in industrial plants.
Verification as per IEC 61000-4-18:
Open circuit
V peak
250V
– 4000V
±10 %
Rise time first peak T1 5ns
±20 %
Oscillating frequency
3 MHz, 10 MHz, 30 MHz ±10 %
Decaying of peak 5
> 50% of peak 1
Decaying of peak 10
< 50% of peak 1
Repetition rate
5000/s
±10 %
Burst duration
3 MHz 50 ms
±20 %
10 MHz 15 ms
±20 %
30 MHz
5 ms
±20 %
Burst period
300 ms
±20 %
Output Impedance
50
Ω
±30 %
Short circuit
Rise time first peak T1 3 MHz < 330 ns
10 MHz < 100 ns
30 MHz
< 33 ns
Osc. current frequency 3 MHz, 10 MHz, 30 MHz ±30 %
Decaying of peak 5
> 25 % of peak 1
Decaying of peak 10
< 25 % of peak 1
I peak
5 A to 80 A
± 20 %
7.1.
Operation
This menu offers different test routines for Damped Oscillatory testing. The menu structure is identical for all, the
3 MHz, 10 MHz and the 30 MHz damped oscillatory.
Page 2
Damped Osc. 3MHz
IEC 61000-4-18
F1 : Quick start
F2 : Standard test routines
F3 : User test routines
F1
F2
F3
F4
F5
F6
F7
F1 Quick start
Easy and fast online-operation of the equipment.
F2 Standard test routines
The operator can select between various preprogrammed test routines as required in different standards
F3 User test routines
The operator can select between various preprogrammed test routines which helps to accelerate testing and
which are very helpful especially during design.