AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
68 / 80
Typical pulses, measured with a calibrated equipment
Fast Damped Oscillatory Wave, 3 MHz, Coaxial output
2000 V on 1000
Ω
Positive
Negative
Fast Damped Oscillatory Wave, 10MHz, Coaxial output
2000 V on 1000
Ω
Positive
Negative
Fast Damped Oscillatory Wave, 30 MHz, Coaxial output
2000 V on 1000
Ω
Positive
Negative
Nominal setting
V Peak
Rise time tr
Decaying [% of Peak 1]
Osc. Freq
[ V ]
[ MHz ]
[ V ]
[ ns ]
Peak 5
Peak 10
[ MHz ]
+2000
3
1963
5.77
66.2
46.5
2.80
-2000
3
-1918
4.88
63.4
42.2
2.82
+2000
10
1976
5.09
54.5
39.3
9.59
-2000
10
-2007
4.91
52.4
38.7
9.73
+2000
30
2022
3.93
57.4
29.0
28.14
-2000
30
-2071
4.30
63.5
24.2
29.89