188
ML-2350A/2351A
2. Maintenance of the Cooler Unit Section
Changing the Ion-Exchange Resin and Replacing
the Ion-Exchanger
The ion-exchange resin in the ion-exchanger has a function to keep high purity by
removing the ion generated by deteriorated secondary cooling water. Change the
ion-exchange resin with a new one within 6 months after use.
For the ion-exchanger of this laser, the cartridge type is adopted. This type can be
repeatedly used if the content (ion-exchange resin) is changed.
Replace the ion-exchanger every 3 years.
⇒
Keep the ion-exchange resin for change in a cool place without being exposed to
direct sunlight if possible. Do not freeze the ion-exchange resin to avoid lowering
its performance.
Item required
Ion-exchanger mounting/removing tool, ion-exchange resin for change (or cartridge),
secondary cooling water (12 liters), Phillips screwdriver, water feed pump, and gloves
(vinyl)
1
Removing the Ion-Exchanger
(1) Open the lid of the cooling water tank and take out the floating panel.
⇒
Keep the panel clean during the exchanging work.
(2) Pump out water in the tank.
(3) Turn the ion-exchanger in the tank to the left with the replacing tool to remove it.
⇒
For changing with a new ion-exchange cartridge (with resin for change), proceed
to Step
3
.
Ion-exchanger
Gloves
Replacing tool
Ion-exchanger
(screw type)
The Laser front (With front door opened)
Содержание ML-2350A-CE
Страница 1: ...AA06OM1173714 18 OPERATION MANUAL YAG LASER WELDER ML CE 2350A 2351A Original instructions...
Страница 19: ...Introduction Part...
Страница 20: ......
Страница 37: ...Installation and Preparation Part...
Страница 38: ......
Страница 46: ......
Страница 57: ...Operating Part...
Страница 58: ......
Страница 118: ......
Страница 128: ......
Страница 150: ......
Страница 174: ......
Страница 179: ...Maintenance Part...
Страница 180: ......
Страница 209: ...Appendixes...
Страница 210: ......