Characteristics and Specifications
9
6000 Series Oscilloscope User’s Guide
357
Logic channels (with MSO option)
Number of channels
16 logic timing channels - labeled D15 - D0
Maximum input frequency
250 MHz
Sample rate
2 GSa/sec one pod, 1 GSa/sec each pod
Standard memory depth
2.5 Mpts one pod, 1.25 Mpts each pod
Vertical resolution
1 bit
Threshold selections
TTL, CMOS, ECL, user-definable (selectable by
pod)
Maximum input voltage
±40 V peak CAT I
Glitch detection
2 ns (min pulse width)
Analog to digital conversion
Vertical resolution
8 bits
Sample rate
DSO6014L: 2 GSa/sec
DSO6054L/6104L: 4 GSa/sec half channel,
2 GSa/sec each channel
Equivalent-time sample rate: 400 GSa/s (when
realtime mode is turned off)
Standard memory depth
8 Mpts per half channel, 4 Mpts per each channel
Time range
5 nsec/div to 50 sec/div (DSO6014L)
1 nsec/div to 50 sec/div (DSO6054L)
500 psec/div to 50 sec/div (DSO6104L)
Acquisition
Acquisition mode
Normal, Peak Detect, Averaging, High Resolution
Peak detection
DSO6014L: 1 nsec peak detect
DSO6054L/6104L: 250 psec peak detect
Averaging
Selectable from 2,4,8,16,32,64... to 65536
High resolution mode
Time base Bits of resolution
< 100 nsec/div
8
500 nsec/div
9
2
µ
sec/div 10
10
µ
sec/div 11
≥
50
µ
sec/div 12
Filter Sinx/x
interpolation
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......