Getting Started
1
6000 Series Oscilloscope User’s Guide
55
By 100 MHz Bandwidth Models
The following active probes use their own power supply.
Therefore, they can be used on all 6000 Series oscilloscopes,
including the 100 MHz bandwidth models.
Many active probes have a 50
Ω
output impedance. The input
impedance of 6000 Series 100 MHz bandwidth models is fixed at
1 M
Ω
. When connecting these probes to 6000 Series 100 MHz
bandwidth models, a 50
Ω
feedthrough terminator (such as
Agilent part number 0960-0301 is required).
Table 11
Active Probes for All 6000 Series Oscilloscopes
Active Probes
Quantity Supported
1141A with 1142A power supply
4
1144A with 1142A power supply
4
1145A with 1142A power supply
2
N2772A with N2773A power supply
4
N2774A with N2775A power supply
4