Characteristics and Specifications
9
6000 Series Oscilloscope User’s Guide
347
Acquisition: logic channels (MSO6000A or MSO-upgraded DSO6000A only)
Sample rate
2 GSa/sec one pod, 1 GSa/sec each pod
Maximum input frequency
250 MHz
Standard memory depth
With oscilloscope channels turned off,
2 Mpts one pod, 1 Mpts each pod
With oscilloscope channels turned on,
625 kpts one pod, 312 kpts each pod
Optional memory depth
With oscilloscope channels turned off,
Option 2ML or 2MH – 2 Mpts one pod, 1 Mpts each pod
Option 8ML or 8MH – 8 Mpts one pod, 4 Mpts each pod
With oscilloscope channels turned on,
Option 2ML or 2MH – 625 kpts one pod, 312 kpts each pod
Option 8ML or 8MH – 2.5 Mpts one pod, 1.25 Mpts each pod
Vertical resolution
1 bit
Glitch detection
2 ns (min pulse width)
Vertical system: oscilloscope channels
Scope channels
MSO/DSO6xx2A: Ch 1 and 2 simultaneous acquisition
MSO/DSO6xx4A: Ch 1, 2, 3 and 4 simultaneous acquisition
AC coupled
MSO/DSO601xA: 3.5 Hz to 100 MHz
MSO/DSO603xA: 3.5 Hz to 300 MHz
MSO/DSO605xA: 3.5 Hz to 500 MHz
MSO/DSO610xA: 3.5 Hz to 1 GHz
Calculated rise time
(= 0.35/bandwidth)
MSO/DSO601xA: 3.5 nsec
MSO/DSO603xA: 1.17 nsec
MSO/DSO605xA: 700 psec
MSO/DSO610xA: 350 psec
Single-shot bandwidth
MSO/DSO601xA: 100 MHz
MSO/DSO603xA: 300 MHz
MSO/DSO605xA: 500 MHz
MSO/DSO610xA: 1 GHz (in half-channel mode, i.e., one channel of channel pair is on)
Range
1
MSO/DSO601xA: 1 mV/div to 5 V/div (1 M
Ω
)
MSO/DSO603xA and MSO/DSO605xA: 2 mV/div to 5 V/div (1 M
Ω
or 50
Ω
)
MSO/DSO610xA: 2 mV/div to 5 V/div (1 M
Ω
), 2 mV/div to 1 V/div (50
Ω
)
Maximum input
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Offset range
±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div; ±75 V on ranges
>200 mV/div
1
1 mV/div is a magnification of 2 mV/div setting for 100 MHz models and 2 mV/div is a magnification of 4 mV/div setting for 300 MHz - 1 GHz models. For
vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting and 32 mV for 2 mV/div sensitivity setting.
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......