Making Measurements
5
6000 Series Oscilloscope User’s Guide
243
Voltage Measurements
Measurement units for each input channel can be set to
Volts or Amps using the channel
Probe Units
softkey. A scale
unit of
U
(undefined) will be displayed for math function 1-2
and for d/dt, and
∫
dt when 1-2 or 1+2 is the selected source
if channel 1 and channel 2 are set to dissimilar units in the
channel
Probe Units
softkey.
The following figure shows the voltage measurement points.
Math Measurements and Units
Only Peak-Peak, Maximum, Minimum, Average, X at Min, and X at Max automatic
measurements may be made on a math FFT function. See “Making time
measurements automatically” for the FFT X at Max and X at Min measurement.
Use the cursors to make other measurements on FFT. All voltage measurements
may be made on other math functions. The resultant units are:
FFT:
1 * 2:
1 – 2:
d/dt:
∫
dt:
dB* (decibels)
V
2
, A
2
or W (Volt-Amp)
V (Volts) or A (Amps)
V/s or A/s (V/second A/second)
Vs or As (V-seconds or A-seconds)
*
When the FFT source is channel 1, 2, 3, or 4, FFT units will be displayed in dBV
when channel units is set to Volts and channel impedance is set to 1 M
Ω
. FFT
units will be displayed in dBm when channel units is set to Volts and channel
impedance is set to 50
Ω
. FFT units will be displayed as dB for all other FFT
sources or when a source channel’s units has been set to Amps.
Maximum
Minimum
Peak-Peak
Amplitude
Top
Base
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......