284
6000 Series Oscilloscope User’s Guide
6
Displaying Data
b
Press the
Source
softkey, and use the Entry knob to select
the channel probing the LIN signal.
c
Repeatedly press and release the
Baud
softkey to specify
the LIN signal baud rate.
d
Repeatedly press and release the
Smpl Pt
softkey to select
the sample point. This controls the point within the bit’s
time where the bit value is captured.
e
Repeatedly press and release the
Standard
softkey to select
the LIN standard (LIN 1.3 or LIN 2.0) of the measured
signal. For LIN 1.2 signals, use the LIN 1.3 setting. The LIN
1.3 setting assumes the signal follows the “Table of Valid
ID Values” as shown in section A.2 of the LIN Specification
dated December 12, 2002. If your signal does not comply
with the table, use the LIN 2.0 setting.
f
Repeatedly press and release the
Sync Break
softkey to
define the minimum number of clocks that define a Sync
Break.
4
Make sure the trigger or threshold levels are set to the middle
of the LIN signal.
•
For analog channels, turn the Trigger Level knob.
•
For digital channels, press the
D15 Thru D0
key and the
Thresholds
softkey to access the threshold level setting
softkeys.
One Bit
60%
70%
80%
Sample Point
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......