Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
147
•
Error Frame
- The oscilloscope will trigger on CAN active
error frames.
•
All Errors
- The oscilloscope will trigger when any form
error or active error is encountered.
•
Acknowledge Error
- The oscilloscope will trigger when the
acknowledge bit is recessive (high).
•
Overload Frame
- The oscilloscope will trigger on CAN
overload frames.
3
If necessary, press the up-arrow softkey to return to the CAN
Trigger menu. Press the
Signals
softkey to enter the CAN
Signals menu.
4
Press the
Source
softkey to select the channel connected to
the CAN signal line.
As you repeatedly press the
Source
softkey (or rotate the
Entry knob), the
CAN
label for the source channel is
automatically set and the channel you select is shown in the
upper-right corner of the display next to “
CAN
”.
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the
D15 Thru D0
key and
select
Thresholds
to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
Baud Rate
Signals
source
Sample
Point
Trigger level
or threshold
Currently selected
signal source
CAN trigger
Return to
previous menu
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......