104
6000 Series Oscilloscope User’s Guide
2
Front-Panel Controls
For a 4-channel oscilloscope, connect BNC tees to the
equal-length cables as shown below. Then connect a
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
Figure 11
User Calibration cable for 4-channel oscilloscope
1
Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
2
Press the
Utility
key, then press the
Service
softkey.
3
Begin the Self Cal by pressing the
Start
User Cal
softkey.
4
When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
User Cal Status
Pressing
Utility
&
Service
&
User Cal Status
displays the summary
results of the previous User Cal, and the status of probe
calibrations for probes that can be calibrated. Note that passive
Longer cable
to TRIG OUT
To Channel 3
To Channel 1
To Channel 4
To Channel 2
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......