![Agilent Technologies InfiniiVision DSO6014L Скачать руководство пользователя страница 195](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-dso6014l/infiniivision-dso6014l_user-manual_2868144195.webp)
194
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
If a more detailed analysis is required, then only one color field
should be selected to be the trigger. You can do this by using the
TV Holdoff
softkey in the trigger More Trigger Menu when the
trigger type is set to
TV
. Press the
TV Holdoff
softkey and use the
Entry knob to adjust the holdoff in half-field increments until
the oscilloscope triggers on only one phase of the color burst.
A quick way to synchronize to the other phase is to briefly
disconnect the signal and then reconnect it. Repeat until the
correct phase is displayed.
When holdoff is adjusted using the
TV Holdoff
softkey and the
Entry knob, the corresponding holdoff time will be displayed in
the
Mode/Coupling
menu.
Figure 19
Triggering on Color Field 1 Alternating with Color Field 3
Содержание InfiniiVision DSO6014L
Страница 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Страница 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Страница 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Страница 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Страница 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Страница 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Страница 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Страница 370: ......